J. Eur. Opt. Society-Rapid Publ. 21, 45( 2025)
Fig
. 6. Results for the simulations of monocytes with cytoplasmic vacuoles and monocytes with cytoplasmic vacuoles and lysosomes. Top panel,( a), an isolevel visualization and a yz slice from a simulated cell, highlighting the positioning of the vacuoles and their spatially-varying RI distribution( Visualization 4). Bottom panel,( b), an isolevel visualization and an xz slice from a simulated cell, where the high RI and numerosity of lysosomes can be appreciated( Visualization 5). EM = External membrane, C = Cytoplasm, N = Nucleus, V = Vacuoles, L = Lysosomes. For each cell, the isolevels’ thresholds for the visualization are set with respect to the maximum value of the simulated RI.
presented in Figure 7, where the hologram was simulated foranexampletakenfromSimulation1, ayeastcellwith a round vacuole inside. An off-axis digital hologram was simulated. The wavelength is 495 nm, the pixel size of the camera is 4.5 lm, the numerical aperture of the objective is 0.95 and the lateral magnification 76.4. The size of the simulated hologram is 5124 5124 pixels. The inclination of the fringes is of 45 ° in the hologram plane. Through this simulation it is possible to visualize also the Fourier spectrum containing the three holographic diffraction orders.
3.2 From the simulated cell to its simulated reconstructed tomogram
In this section, example use-cases for the presented cell models are reported. Concerning the nomenclature, we herein refer to the 3D RI cellular model proposed as“ simulated cell”, which is the high-fidelity representation of the 3D RI distribution of the sample of interest. Instead, the tomogram is the simulated reconstruction of the 3D RI distribution, after the cell has been probed through an arbitrary optical system and reconstructed through an arbitrary inversion algorithm.
The tomogram hence embeds the effects of the optical setup, noise, and the specific algorithms exploited for each reconstruction step, including the tomographic inversion algorithms. All of these aspects determine a difference between the“ simulated cell” and its relative“ tomogram”, which is a degraded version of the cell after the passage through a non-ideal holo-tomographic measurement-reconstruction system. It is worth pointing out again that the aim of our work was to propose a high-fidelity ground truth cellular model which could be used to tune different algorithms relative to the holo-tomographic acquisition-reconstruction pipeline. Depending on the specific holographic setup or tomographic paradigm employed, the demonstrative and simplified use-cases discussed in this section can be modified to suit the one of interest, for example embedding the measured PSF of the setup or exploiting different tomographic solvers to reconstruct the tomogram.
In the reported examples, we consider the effect of an isotropic system PSF operating as a 3D low-pass filter on the cell. This is reasonable in all the cases of in-flow holotomography where isotropic resolution is achievable. Then, we consider the effects of speckle noise as well as the use of the tomographic inversion algorithms. Two different classes