J. Eur. Opt. Society-Rapid Publ. 21, 45( 2025) 151
Fig. 7. Left: Numerical digital hologram obtained starting from a simulated yeast cell with a vacuole( Simulation 1). The interference fringes are highlighted in the yellow box. Right: corresponding simulation of the amplitude Fourier spectrum, in which the holographic diffraction orders are clearly visible.
of inversion algorithms are considered here, namely the filtered backprojection [ 44 ] and the filtered backpropagation with Rytov approximation [ 45 ].
After the ground-truth 3D RI distributions are simulated, the system PSF, speckle noise, and the influence of the employed inversion algorithm have to be considered. Here, a simplified general method is considered, which allows to simulate and test each step of the holo-tomographic acquisition-reconstruction pipeline [ 3, 46, 47 ]. This is summarized in the sketch of Figure 8a. The starting point is the cell simulation we propose in this work. Convoluting this model with the PSF of the optical system in use, the simulated tomogram model is obtained. Reprojecting this model along the available tomographic angles allows simulating the QPMs, which can be eventually corrupted by noise [ 48 ]. Finally, composing these QPMs with a suitable tomographic solver [ 48 ] allows computing the reconstructed, simulated tomogram. Figure 8b represents each step of the considered pipeline applied to our in-flow tomographic flow cytometry optical setup for a sample simulated cell.
1. A RI distribution of interest, C, of size n n n is simulated according to the presented numerical models. Figure 8b1 sketches the central slice of C.
2. The PSF of the optical system is modeled through a 3D Gaussian kernel G of size 5 5 5 pixels and standard deviation 2. These values where experimentally chosen for this specific example to obtain a loss in the resolution visually similar to the one observed in real reconstructed tomograms. Figure 8b2 depicts the central slice of the Gaussian kernel.
3. The simulated tomogram T of size n n n is obtained as the convolution between the modeled PSF and the RI distribution C. In formulas, T = C b G, where b is the convolution operator. Figure 8( b3) sketches the central slice of T.
4. The QPMs are obtained by reprojecting T along the available projection angles. In formulas Q = w { T, theta }, where w {...} is the operator modelling the projection, theta is the set of projection angles with size 1 t and Q is the stack of resulting QPMs of size n n t. In our case, the operator w {...} is obtained through the MATLAB â radon function [ 49 ]. Figure 8b4 depicts one of the obtained QPMs.
5. The noisy QPMs, QPM noise, are obtained by adding a suitable noise distribution N over the uncorrupted QPMs. In our case, the noise is modeled as a distribution of multiplicative speckle noise with mean 1 and variance 5e – 6. This was made resorting to the imnoise MATLAB â function with“ speckle” specification [ 50 ].
6. The simulated reconstructed tomogram T rec is obtained by inverting the operator w {...}, composing all the available projections and the associate projections angles with a suitable physical criterion. In our case, this step is performed resorting to the MATLAB â iradon routine in the first example [ 44 ] and using a custom filtered backpropagation routine using the Rytov approximation in the second example [ 45 ]. In formulas, T rec ¼ w �1 fQPM noise; thetag. Summing up, the simulated reconstructed tomogram is obtainable from the simulated cell as:
T rec ¼ w �1 w CbG; theta þ N; theta ð1Þ
Figure
8b6 depicts the central slice of the simulated reconstructed tomogram T rec.
Figure 9 depicts the application of this pipeline to the cells simulated and discussed in this work. The top row shows an yz slice from the simulated cells in Figures 3, 5 and 6a, the central row shows the resulting simulated tomogram reconstruction obtained through the filtered backprojection algorithm, and the bottom row shows the reconstructions obtained through the filtered backpropagation with Rytov approximation. Each step of the pipeline is important and can be flexibly used to tailor suitable algorithms of interest according to the specific needs of the user and the features of the employed optical system. For example, we can investigate the effects of limited projection angles and the impact of errors in projection angle recovery, particularly in in-flow setups. In addition, denoising procedures can be experimented to study their effects on QPMs or even on the tomogram.