DynaNail® TTC Fusion System - Instructions For Use | MedShape DynaNail® TTC Fusion System - Instructions For Us | Page 9

*Alterna)ve Technique* If the surgeon would like to bypass manual compression, steps 6-8 may be replaced with the following: 6b. With the Anvil aEached to the Deployment Frame, strike the Anvil with a mallet to compress the arthrodesis site. 7b. Drill a hole into the distal )bial oblong slot of the DynaNail Implant and insert a 5mm Cor)cal Headed Screw 1 . 8b. (OPTIONAL) Drill a hole into one of the proximal )bial screw holes of the DynaNail Implant and insert a 5mm Cor)cal Headed Screw. 9. Release the DynaNail Implant from the Deployment Frame by unscrewing the Connec@on Screw. 10. Place the End Cap into the distal end of the DynaNail Implant and close incision per surgeon preference. 1 The Deployment Frame, Guide Sleeve and appropriate Drill Guide may be used to target all drills/screws except for the most proximal hole on the 26 cm and 30 cm length DynaNail implants. Refer to the Surgical Technique Guide for further instruc@ons. 12. HOW SUPPLIED The DynaNail TTC Fusion Nail, DynaNail End Cap, and DynaNail Screws are provided sterile for single use only. Carefully inspect sterile packaging for damage prior to use. If the sterile packaging is found to be damaged or open, do not use the device or aZempt to resterilize. Call your MedShape sales representa@ve for a replacement. The DynaNail Deployment Frame and the Ancillary Surgical Instruments are provided NON-STERILE. The DynaNail Deployment Frame and each Ancillary Surgical Instrument must be properly cleaned and sterilized prior to first use and before each subsequent use in accordance with the guidelines provided herein. 13. CLEANING AND STERILIZATION PROCEDURES (DEPLOYMENT FRAME AND ANCILLARY SURGICAL INSTRUMENTATION ONLY) These general recommenda@ons are provided as a guide only. All cleaning and steriliza@on of the DynaNail Deployment Frame and Ancillary Surgical Instruments should be carried out by proficient personnel following appropriate healthcare facility procedures. The DynaNail Deployment Frame and the Ancillary Surgical Instruments are designed to be cleaned and steam sterilized following the instruc@ons below.