194
J. Eur. Opt. Society-Rapid Publ. 21, 50( 2025)
Fig. 8. SXR measurements and RCWA model for( a) the periodic multilayer grating N = 6 and( b) the aperiodic multilayer grating in classical mounting.
Fig
. 9. SXR measurements for the periodic multilayer grating N = 6 in conical mounting( X 0 °). The symbols with dashed line correspond to measurement after a 180 ° rotation of the sample( X = X 0 + 180 °).
theoretically we expect the same efficiency for both orders due to the symmetry of conical diffraction.
The only possible explanation for this difference in peak efficiency between + 1 and �1 order in conical mounting is a loss of symmetry, either in the experimental set-up or in the sample.
During the experiment reported in Figure 9, the sides of the rectangular grating substrate were aligned with the beam, but an offset angle may exist between the direction of the grooves and the side of the substrate. Let’ s consider that the initial measurements were done with an azimuthal angle X 0 between the incident beam and the grooves. From the measured out-of-plane positions of the diffracted orders, we determined that the groove direction was aligned with the incident beam within 0.2 °. Simulation confirms that X 0 values in this range cannot explain the difference in peak efficiency observed between + 1 and �1 orders. Figure 10 shows the + 1 and �1 order efficiency spectra simulated as a function of X. We can notice that close to conical diffraction( X = 0 °) the position of both diffracted orders is sensitive to a small change in X and that they vary in opposite direction. However, the peak efficiency for both orders vary only slightly with X. Consequently, the misalignment of the grating groove with the direction of theincidentbeamisnotresponsibleforthedifferencein peak efficiency observed in Figure 9.
Furthermore, we re-measured the grating efficiency after a 180 ° rotation of the sample( X = X 0 + 180 °). If the efficiency difference observed between the + 1 and �1 orders was due to an asymmetry in experimental setup, we would expect to obtain the same result for the 2 configurations( X 0 and X 0 + 180 °) because the tilt of the groove will always be in the same direction. However, we measured exactly the opposite effect: After the rotation of 180 ° the �1 order provides higher efficiency than the + 1 order( Fig. 9). This means that the loss of symmetry is inherent to the sample itself.
In order to account for the asymmetry of the ML grating, we implemented a model with different slopes, respectively a 1 and a 2, ontheleftandrightedgesofthe grooves( see inset in Fig. 11a). The grating parameters for these models are given in Table 3. Inthesamewayasthe previous model( see Fig. 8), the DW factor was not included in this modeling, as shown in Figure 11, theasymmetric groove model agrees well with experimental data for the 3 diffraction orders in the case of the periodic ML grating both in classical and conical mountings( Figs. 11a and 11b respectively). Furthermore, we used in this model the newly available optical constant for Al [ 32 ], which improves slightly the agreement in the spectral shape and near the Al L 2, 3 absorption edge. The asymmetric groove model also provides better agreement with experimental data in the case of the aperiodic ML grating in classical and conical mountings( Figs. 11c and 11d respectively). Nevertheless, a discrepancy remains for 0 order and �1 order efficiencies. As explained earlier, this can be due to the defects observed by TEM for the first deposited layers of this aperiodic ML which are not taken into account in the model.