T & C and acceptance shall be carried out for all newly installation of active medical device before it is placed into clinical service and shall include but not limited to other safety aspect , visual inspection , EST and performance test .
Section 6.4 :
The supplier shall carry out testing & commissioning and acceptance in the presence of Biomedical Engineering Services and the user . The results of the acceptance testing shall be verified by the Biomedical Engineering Services . Results shall be documented and handed over together with all relevant certificates to the Biomedical Engineering Services .
“ Definition : Biomedical Engineering Services - Any in-house or third-party personnel or organisation or manufacturer / AR recognised by an authorised agency responsible to carry out , implement and / or manage Biomedical Engineering activities ”