8
SURGICAL TECHNIQUE
SUBPECTORAL BICEPS
Implant Positioning
4a
4b
4a
Position the Implant Sheath in the tunnel so that the compressed profile is adjacent to the tendon.
Advance the Sheath into the tunnel. If necessary, mallet the metal Strike Plate on the Gun Knob to
assist in advancing the Sheath into the bone tunnel. Do not twist or rotate the Deployment Gun while
advancing the Implant into the bone tunnel.
Pearl #2. Ensure the Sheath is inserted until it is flush with the cortex of the bone.
Gun Body Rotation!
5a
5b
If desired, push forward on the blue Knob Lock to allow free rotation of the Gun Body. The Gun Body
can now rotate to the desired deployment position for surgeon comfort without disturbing Implant
alignment in the tunnel. Push back on the Knob Lock to re-engage with the Gun Knob, preventing
further rotation of the Gun Body.!