7
SURGICAL TECHNIQUE
SUBPECTORAL BICEPS
Tendon Handling, Tendon Fork
3a
3b
Option 1: Using the Tendon Fork, push the suture-tendon complex into the bone tunnel. Verify the
length of tendon in the socket by reading the lasermarks on the Tendon Fork.
Tendon Handling, Suture Lasso
3c
3d
Option 2: Alternatively, the Suture Lasso may be used to aid in tendon insertion. Use the Suture
Lasso to pull one suture thread through the Sheath. The second suture thread should run along the
outside of the Implant Sheath. Adjust the tendon so that it rests in the compressed profile of the
Sheath. Cleat the threads on the Suture Cleat on the back of the Gun and insert the Implant and
tendon into the bone tunnel. Remember to uncleat the sutures before releasing the Gun from the
Implant.
Note: While the Suture Lasso is a viable option, MedShape recommends using the Tendon
Fork for easier tendon management.