The Journal of ExtraCorporeal Technology No 57-1 | Page 52

46 D . Rusnak et al .: J Extra Corpor Technol 2025 , 57 , 42 – 49
Figure
2 . Plasma Leakage on Polypropylene Oxygenator .
encountered , including thrombosis and high transmembrane pressures , each oxygenator had varying durations of use . EUROSETS oxygenators were notably durable , with the third EUROSETS model ( EUROSETS nr 3 ) lasting the longest at 88 days . In contrast , the Terumo Fx15 models had shorter durations , with some lasting only 1 – 2 days , primarily due to plasma leakage , which were recurrent issues with polypropylene-based oxygenators . Additionally , plasma leakage was particularly observed in the Terumo Fx15 oxygenators , necessitating more frequent replacements ( Figure 2 ). The high LDH levels and elevated free Hb , particularly with the Quadrox-i and Terumo Fx15 models , indicate the hemolytic stress and ongoing coagulation challenges faced over the course of treatment . Continuous anticoagulation was maintained throughout the ECMO support , although three Terumo oxygenators were operated without anticoagulation to manage hemorrhage risks . The tables summarize the performance parameters and laboratory parameters observed across each oxygenator type during the treatment . Parameters such as Hb , PaO 2 post-ox , PaCO 2 post-ox , and FiO 2 show the adaptation to each oxygenator ’ s functionality , with fluctuations reflecting the technical limitations and biological responses ( Table 1 ). Table 2 summarizes the highest and lowest parameters from the scientific evidence for each oxygenator models presented as mean and deviation standard values . Figure 3 presents the graphical trend of the values in the different Oxygenator models .
Discussion
This case highlights the profound challenges of managing prolonged ECMO support in a resource-limited , conflictaffected environment . Prolonged ECMO support for 259 days is an exceptional scenario , particularly in a war-torn setting
Table 2 . The highest and lowest parameters for each oxygenator models presented as mean and deviation standard values .
Lowest Pressure Drop ( DP ) ( Mean ± SD )
( mmHg )
Highest Pressure Drop ( DP ) ( Mean ± SD )
( mmHg )
Lowest Blood Flow Rate ( Mean ± SD )
( L / min )
Highest Blood Flow Rate ( Mean ± SD )
( L / min )
Lowest Air Flow
( Mean ± SD ) ( L / min )
Highest Air Flow
( Mean ± SD ) ( L / min )
Lowest pCO2 post-ox
( Mean ± SD ) ( mmHg )
Highest pCO2 post-ox
( Mean ± SD ) ( mmHg )
Lowest pO2 post-ox
( Mean ± SD ) ( mmHg )
Highest
pO2 post-ox ( Mean ± SD )
( mmHg )
Duration ( Days )
Oxygenator Model
EUROSETS nr1
10
419 ± 34
215 ± 20
32 ± 5
52 ± 6
2.0 ± 0.3
2.5 ± 0.2
3.5 ± 0.5
3.2 ± 0.4
61 ± 5
67 ± 6
EUROSETS nr2
32
326 ± 19
207 ± 18
26.4 ± 4
40.2 ± 4
0.8 ± 0.1
4.0 ± 0.3
2.41 ± 0.3
1.7 ± 0.2
42 ± 4
75 ± 5
EUROSETS nr3
88
361 ± 27
216 ± 25
31 ± 4
39 ± 4
0.8 ± 0.1
4.0 ± 0.3
2.33 ± 0.3
1.9 ± 0.2
40 ± 4
75 ± 5
EUROSETS nr4
26
351 ± 19
224 ± 30
32 ± 5
38 ± 5
0.8 ± 0.1
5.0 ± 0.5
2.23 ± 0.2
0.5 ± 0.1
41 ± 5
170 ± 10
Maquet PLS
14
324 ± 20
60 ± 5
28 ± 3
42 ± 4
1.5 ± 0.2
8.0 ± 0.5
2.2 ± 0.3
2.0 ± 0.2
25 ± 3
75 ± 5
Paragon PMP
78
473 ± 45
230 ± 35
23 ± 3
37 ± 4
1.0 ± 0.2
2.0 ± 0.2
2.2 ± 0.3
2.0 ± 0.2
35 ± 4
57 ± 4
Terumo Fx15 nr1
5
287 ± 30
125 ± 15
30 ± 4
42 ± 4
2.0 ± 0.3
4.0 ± 0.3
2.31 ± 0.3
2.0 ± 0.3
31 ± 4
67 ± 5
Terumo Fx15 nr2
1
346 ± 40
138 ± 18
30.3 ± 3
34 ± 5
2.0 ± 0.3
3.0 ± 0.2
2.3 ± 0.2
2.2 ± 0.3
35 ± 5
40 ± 3
Terumo Fx15 nr3
3
321 ± 25
115 ± 14
32 ± 5
40 ± 4
2.0 ± 0.2
4.0 ± 0.3
2.4 ± 0.2
2.3 ± 0.2
32 ± 4
54 ± 5
Terumo Fx15 nr4
2
399 ± 30
98 ± 10
31 ± 4
38 ± 4
2.0 ± 0.3
4.0 ± 0.2
2.3 ± 0.2
2.2 ± 0.3
40 ± 5
41 ± 4
Quadrox-i
1
240 ± 35
121 ± 20
31 ± 3
40 ± 5
2.0 ± 0.3
4.0 ± 0.3
2.7 ± 0.3
2.3 ± 0.2
35 ± 4
50 ± 5