46 D . Rusnak et al .: J Extra Corpor Technol 2025 , 57 , 42 – 49
Figure
2 . Plasma Leakage on Polypropylene Oxygenator .
encountered , including thrombosis and high transmembrane pressures , each oxygenator had varying durations of use . EUROSETS oxygenators were notably durable , with the third EUROSETS model ( EUROSETS nr 3 ) lasting the longest at 88 days . In contrast , the Terumo Fx15 models had shorter durations , with some lasting only 1 – 2 days , primarily due to plasma leakage , which were recurrent issues with polypropylene-based oxygenators . Additionally , plasma leakage was particularly observed in the Terumo Fx15 oxygenators , necessitating more frequent replacements ( Figure 2 ). The high LDH levels and elevated free Hb , particularly with the Quadrox-i and Terumo Fx15 models , indicate the hemolytic stress and ongoing coagulation challenges faced over the course of treatment . Continuous anticoagulation was maintained throughout the ECMO support , although three Terumo oxygenators were operated without anticoagulation to manage hemorrhage risks . The tables summarize the performance parameters and laboratory parameters observed across each oxygenator type during the treatment . Parameters such as Hb , PaO 2 post-ox , PaCO 2 post-ox , and FiO 2 show the adaptation to each oxygenator ’ s functionality , with fluctuations reflecting the technical limitations and biological responses ( Table 1 ). Table 2 summarizes the highest and lowest parameters from the scientific evidence for each oxygenator models presented as mean and deviation standard values . Figure 3 presents the graphical trend of the values in the different Oxygenator models .
Discussion
This case highlights the profound challenges of managing prolonged ECMO support in a resource-limited , conflictaffected environment . Prolonged ECMO support for 259 days is an exceptional scenario , particularly in a war-torn setting
Table 2 . The highest and lowest parameters for each oxygenator models presented as mean and deviation standard values .
Lowest Pressure Drop ( DP ) ( Mean ± SD )
( mmHg )
Highest Pressure Drop ( DP ) ( Mean ± SD )
( mmHg )
Lowest Blood Flow Rate ( Mean ± SD )
( L / min )
Highest Blood Flow Rate ( Mean ± SD )
( L / min )
Lowest Air Flow
( Mean ± SD ) ( L / min )
Highest Air Flow
( Mean ± SD ) ( L / min )
Lowest pCO2 post-ox
( Mean ± SD ) ( mmHg )
Highest pCO2 post-ox
( Mean ± SD ) ( mmHg )
Lowest pO2 post-ox
( Mean ± SD ) ( mmHg )
Highest
pO2 post-ox ( Mean ± SD )
( mmHg )
Duration ( Days )
Oxygenator Model
EUROSETS nr1 |
10 |
419 ± 34 |
215 ± 20 |
32 ± 5 |
52 ± 6 |
2.0 ± 0.3 |
2.5 ± 0.2 |
3.5 ± 0.5 |
3.2 ± 0.4 |
61 ± 5 |
67 ± 6 |
EUROSETS nr2 |
32 |
326 ± 19 |
207 ± 18 |
26.4 ± 4 |
40.2 ± 4 |
0.8 ± 0.1 |
4.0 ± 0.3 |
2.41 ± 0.3 |
1.7 ± 0.2 |
42 ± 4 |
75 ± 5 |
EUROSETS nr3 |
88 |
361 ± 27 |
216 ± 25 |
31 ± 4 |
39 ± 4 |
0.8 ± 0.1 |
4.0 ± 0.3 |
2.33 ± 0.3 |
1.9 ± 0.2 |
40 ± 4 |
75 ± 5 |
EUROSETS nr4 |
26 |
351 ± 19 |
224 ± 30 |
32 ± 5 |
38 ± 5 |
0.8 ± 0.1 |
5.0 ± 0.5 |
2.23 ± 0.2 |
0.5 ± 0.1 |
41 ± 5 |
170 ± 10 |
Maquet PLS |
14 |
324 ± 20 |
60 ± 5 |
28 ± 3 |
42 ± 4 |
1.5 ± 0.2 |
8.0 ± 0.5 |
2.2 ± 0.3 |
2.0 ± 0.2 |
25 ± 3 |
75 ± 5 |
Paragon PMP |
78 |
473 ± 45 |
230 ± 35 |
23 ± 3 |
37 ± 4 |
1.0 ± 0.2 |
2.0 ± 0.2 |
2.2 ± 0.3 |
2.0 ± 0.2 |
35 ± 4 |
57 ± 4 |
Terumo Fx15 nr1 |
5 |
287 ± 30 |
125 ± 15 |
30 ± 4 |
42 ± 4 |
2.0 ± 0.3 |
4.0 ± 0.3 |
2.31 ± 0.3 |
2.0 ± 0.3 |
31 ± 4 |
67 ± 5 |
Terumo Fx15 nr2 |
1 |
346 ± 40 |
138 ± 18 |
30.3 ± 3 |
34 ± 5 |
2.0 ± 0.3 |
3.0 ± 0.2 |
2.3 ± 0.2 |
2.2 ± 0.3 |
35 ± 5 |
40 ± 3 |
Terumo Fx15 nr3 |
3 |
321 ± 25 |
115 ± 14 |
32 ± 5 |
40 ± 4 |
2.0 ± 0.2 |
4.0 ± 0.3 |
2.4 ± 0.2 |
2.3 ± 0.2 |
32 ± 4 |
54 ± 5 |
Terumo Fx15 nr4 |
2 |
399 ± 30 |
98 ± 10 |
31 ± 4 |
38 ± 4 |
2.0 ± 0.3 |
4.0 ± 0.2 |
2.3 ± 0.2 |
2.2 ± 0.3 |
40 ± 5 |
41 ± 4 |
Quadrox-i |
1 |
240 ± 35 |
121 ± 20 |
31 ± 3 |
40 ± 5 |
2.0 ± 0.3 |
4.0 ± 0.3 |
2.7 ± 0.3 |
2.3 ± 0.2 |
35 ± 4 |
50 ± 5 |