T . Matsumoto et al .: J Extra Corpor Technol 2023 , 55 , 167 – 174 171
Table 2 . Simultaneous reproducibility with test liquids 1 – 7 in Experiment 1
|
Measurement method ( n = 10 ) |
Mean COP [ mmHg ] |
SD |
CV [%] |
Test liquid 1 ( Hct : 20.2 %) |
Test method |
5.77 |
3.50 10 �2 |
0.6 |
|
Colloid osmometer |
5.84 |
4.90 10 �2 |
0.8 |
Test liquid 2 ( Hct : 22.6 %) |
Test method |
7.09 |
1.28 10 �1 |
1.8 |
|
Colloid osmometer |
7.08 |
1.10 10 �1 |
1.5 |
Test liquid 3 ( Hct : 23.8 %) |
Test method |
8.23 |
1.49 10 �1 |
1.8 |
|
Colloid osmometer |
8.11 |
1.00 10 �1 |
1.3 |
Test liquid 4 ( Hct : 26.0 %) |
Test method |
9.77 |
8.35 10 �2 |
0.9 |
|
Colloid osmometer |
9.6 |
7.70 10 �2 |
0.8 |
Test liquid 5 ( Hct : 28.0 %) |
Test method |
11.65 |
9.45 10 �2 |
0.8 |
|
Colloid osmometer |
11.48 |
1.10 10 �1 |
0.9 |
Test liquid 6 ( Hct : 31.7 %) |
Test method |
15.51 |
4.07 10 �1 |
2.6 |
|
Colloid osmometer |
14.81 |
1.80 10 �1 |
1.2 |
Test liquid 7 ( Hct : 34.8 %) |
Test method |
19.69 |
3.12 10 �1 |
1.6 |
|
Colloid osmometer |
18.34 |
6.60 10 �2 |
0.4 |
Measurements were performed 10 times for each test liquid and each measurement method , and CVs (%) were calculated . For the COP measurements using the Test Method , all CVs were in the range of 1.44 ± 0.66 %. For measurements obtained using the colloid osmometer , all CVs were in the range of 0.99 % ± 0.34 %. Abbreviations : COP , colloid osmometer ; SD , standard deviation ; CV , coefficient of variation ; Hct , hematocrit value .
Correlation
COP as measured by the test method and the colloid osmometer demonstrated a strong correlation ( Pearson product-moment correlation coefficient R 2 = 0.997 ; p < 0.01 ). The formula for the regression line is as follows ( Figure 5 ):
y ¼ 0:5351251 þ 0:9813299 x : ð2Þ
Measurement of time
Although the standby time was 3 min , solvent transfer was completed in under 1 min for test liquid 8 . Solvent transfer time increased as the concentration of the test liquid increased . With test liquid 10 , moderate solvent migration appeared to occur after 1 min , but no solvent migration could be observed after 2 min had passed .
Discussion
Evaluation of measurement accuracy ( simultaneous reproducibility )
Simultaneous reproducibility was used to evaluate the accuracy of the measured values . Each measurement was within 3 %, which indicates high reproducibility ( Tables 2 and 3 ). The presence of little variability indicates that there may not be many factors contributing to variations in the measurements , even when the person obtaining the measurement changes . This suggests that this measurement method is highly reliable . The absolute value of the colloid osmometer measurement error is ± 5 mmHg . It has an error rate of 10 % in calculating albumin concentrations [ 19 ]. Because the systematic error rate of the test method was unknown , we could not accurately capture all the errors . However , the measurement result was similar to that of the colloid osmometer , which has a measurement error rate of
Figure 4 . Correlation between COP measurements from the colloid osmometer and by the test method used in Experiment 1 . n = 70 ; Pearson ’ s correlation : p < 0.01 ; R 2 = 0.998 . COP , colloid osmotic pressure . y = �0.815074 + 1.1085785 x .
less than 3 % CV . Thus , the systemic error rate of the test method is expected to be comparable to that of the colloid osmometer .
Over extended periods of use , the measurement accuracy may be impacted by the performance of the ultrafiltration membranes , which are not constant . The concentration polarization model can explain the filtration performance of semipermeable membranes ; substances that cannot pass through the membrane concentrate on the membrane surface increasing transmembrane resistance and decreasing filtration performance . Additionally , fouling is caused by the adhesion of proteins to the surface and interior of the membrane , which similarly increases the membrane passage resistance and reduces the permeability and solute permeability [ 20 – 22 ]. These factors will be affected by the filtered flow rate , blood flow rate , blood concentration , and elapsed time . During our experiment , the ultrafiltration membrane was loaded using variable rates of blood flow . In addition , the blood concentration and elapsed time were varied .