GAS FILTER CORRELATION
KEY APPLICATIONS
• Continuous emissions monitoring
• Ethylene , chlorine and TDI production processes
• HyCO process control
KEY BENEFITS
• Industry-leading lower detection limit
• Fast response and rapid recovery
• Non-depleting sensor – long lifespan
IDEAL FOR
Sensitive , parts-per-million measurements of O 2 , for example in impurity monitoring for UHP semiconductor gases .
WORKS WITH
Single-beam , single-wave Infrared sensing to provide real-time process analysis for a range of industrial applications .
LIMITATIONS
Only gases with infrared lines can be measured by this technology , so it is not suitable for noble gases , or single element diatomic molecules such as N 2 or O 2 .
USED IN
SERVOTOUGH SpectraExact 2500 SERVOPRO MultiExact 4100
SERVOPRO MultiExact 4200
SERVOPRO 4900 Multigas
P81