P & S APPLICATION
MEDICAL GASES
Gases for medical treatment are regulated like medicinal drugs . These regulations – typically covered in a publication called a Pharmacopeia – specify production and validation methods , the acceptable purity level , and official measurement records .
For example , under European Pharmacopeia ( EP ) rules , medical oxygen ( O 2 ) requires an assay measurement to ensure O 2 purity is better than 99.5 %, and impurity measurements of carbon monoxide ( CO ) and carbon dioxide ( CO 2 ). The impurities must be less than 5 parts per million ( ppm ) of CO and less than 300ppm of CO 2 .
Our high-performance solutions and technologies deliver the measurements required to meet US and European Pharmacopeia concentration limits for medical gas quality using industryapproved sensing techniques .
SERVOPRO MultiExact 4100
P30