SACOR BUSINESS GROUP-2019 COMPANY CATALOGUE SACOR BUSINESS GROUP 2019 COMPANY CATALOGUE | Page 41

GENERAL FEATURES: Transmitter type Battery life Accuracy Battery powered - 2 x D Cell 3,6 V * Lithium battery pack up to 10 years 0,2 % +/- 2 mm/s - insertion sensors 2% of rate +/- 2mm/s Ambient: -20 ... +60 C° (-4 ... +140 F) Temperature Media 0 ... 70 C° (32 ... +158 F) Storage -40 ... +70 C° (-22 ... +158 F) Enclosure Cable entries Custody transfer Policarbonate case with aluminum bottom on compact vertical version. IP 67 / 68. Remote wall mount braket in carbon steel zinch plated 4X PG9 Glands I/O - 2X PG20 Glands junction box in remote version Type approved OIML R49-1 2013 / EN 14154 MID EN-ISO 4064 - Certificate n. T10713 Conformity EMC: EN 61000-6-3, EN 61000-6-2, EN 61326-1 Sensor type Full bore up to DN 600 - Insertion meters Flow velocity range Sampling rate Installation Digital filters 0,015 m/s up to 10 m/s Standard mode 1 / 5 Hz up to 1 / 60 Hz (default 1 / 15 Hz) max 3,125 Hz Integral (compact) or remote with factory mounted sensor cable in 5 m (16,4 ft) up to 30 m (98,4 ft) Damping - cut-off (0,05 m/s default) - bypass - peak cut LCD display - Index, menu, and symbols icons for dedicated information Display and keys 4 Push buttons to access all functions Totalizer informations can be displayed with 5 decimal digits Displayed informations Flow Units Outputs Communication Datalogging Live flowrate Total positive totalizer (T+) Total negative totalizer (T-) Partial positive totalizer (P+) Partial negative totalizer (P-) Time & date Converter temperature Process pressure and temperature (if available) Parameters corresponding code and value m, m 3 , l, ML, ft 3 , GAL 2 pulses passive outputs (MOS), individual galvanically isolated - clean contact Maximum load +/- 35V DC, 100 mA short circuit protected Integrated Euromag IrComm interface 100,000 lines of data with a frequency of log between 1 minute and 120 minutes (default 15 minutes) GSM/GPRS Euromag Module Add on modules Pressure (1) and temperature (2) Energy metering ready Totalizers 4 (2 positive and 2 negative) Date and time Data protection Alarms and status Self diagnostic Yes Password available, automatic firmware check and recover during the update Status icon displayed and alarm logged in the datalogger Alarms available: • excitation failure • empty pipe on the 4th electrode • empty pipe on the measuring electrodes • high temperature • • • high voltage supply pulse overlapped wet electronic board External verification Field verificator available for calibration verification and electronic status Software for communication and programming Commissioning (equal settings of meters) - Data print for documentation - Data export (CSV file) - Firmware update - Read instant flowrate - Read and write all non volatile parameters - Download internal datalogger - View instrument event logger * Lithium batteries are subject to special transportation regulations according to United Nations “Regulation of Dangerous Good”, UN 3090 and UN 3091. Special trasnport documentation is required to observe these regulations. This may influence both trasnport time and cost Product Catalogue 35