GENERAL FEATURES:
Transmitter type
Battery life
Accuracy
Battery powered - 2 x D Cell 3,6 V *
Lithium battery pack up to 10 years
0,2 % +/- 2 mm/s - insertion sensors 2% of rate +/- 2mm/s
Ambient: -20 ... +60 C° (-4 ... +140 F)
Temperature
Media 0 ... 70 C° (32 ... +158 F)
Storage -40 ... +70 C° (-22 ... +158 F)
Enclosure
Cable entries
Custody transfer
Policarbonate case with aluminum bottom on compact vertical version.
IP 67 / 68. Remote wall mount braket in carbon steel zinch plated
4X PG9 Glands I/O - 2X PG20 Glands junction box in remote version
Type approved OIML R49-1 2013 / EN 14154 MID EN-ISO 4064 - Certificate n. T10713
Conformity EMC: EN 61000-6-3, EN 61000-6-2, EN 61326-1
Sensor type Full bore up to DN 600 - Insertion meters
Flow velocity range
Sampling rate
Installation
Digital filters
0,015 m/s up to 10 m/s
Standard mode 1 / 5 Hz up to 1 / 60 Hz (default 1 / 15 Hz) max 3,125 Hz
Integral (compact) or remote with factory mounted sensor cable in 5 m (16,4 ft) up to 30 m (98,4 ft)
Damping - cut-off (0,05 m/s default) - bypass - peak cut
LCD display - Index, menu, and symbols icons for dedicated information
Display and keys
4 Push buttons to access all functions
Totalizer informations can be displayed with 5 decimal digits
Displayed informations
Flow Units
Outputs
Communication
Datalogging
Live flowrate
Total positive totalizer (T+)
Total negative totalizer (T-)
Partial positive totalizer (P+)
Partial negative totalizer (P-)
Time & date
Converter temperature
Process pressure and temperature (if available)
Parameters corresponding code and value
m, m 3 , l, ML, ft 3 , GAL
2 pulses passive outputs (MOS), individual galvanically isolated - clean contact
Maximum load +/- 35V DC, 100 mA short circuit protected
Integrated Euromag IrComm interface
100,000 lines of data with a frequency of log between 1 minute and 120 minutes (default 15 minutes)
GSM/GPRS Euromag Module
Add on modules
Pressure (1) and temperature (2)
Energy metering ready
Totalizers
4 (2 positive and 2 negative)
Date and time
Data protection
Alarms and status
Self diagnostic
Yes
Password available, automatic firmware check and recover during the update
Status icon displayed and alarm logged in the datalogger
Alarms available:
•
excitation failure
•
empty pipe on the 4th electrode
•
empty pipe on the measuring electrodes
•
high temperature
•
•
•
high voltage supply
pulse overlapped
wet electronic board
External verification Field verificator available for calibration verification and electronic status
Software for
communication and
programming Commissioning (equal settings of meters) - Data print for documentation - Data export
(CSV file) - Firmware update - Read instant flowrate - Read and write all non volatile
parameters - Download internal datalogger - View instrument event logger