M . EL Fahssi et al .: Radioprotection 2024 , 59 ( 1 ), 13 – 18 15
Table 3 . The mean of CTDIvol , DLP , and effective dose for Brain C� , Chest C� and CAP Cþ .
Brain C�
Chest C�
Parameters
Mean
Range
SD
Mean
Range
SD
Mean
Range
SD
CTDI vol
53,80
20,73 – 78,04
14,43
9,64
5,48 – 15,83
3.29
7,17
4,55 – 10,33
1.80
DLP
1131,44
607,07 – 1669,28
339,60
367,15
205,35 – 643,10
137.6
465,73
219,09 – 666,49
128.94
Effective dose
2,37
1,27 – 3,50
0,71
6,50
3,63 – 11,38
2.43
7,13
3,29 – 10,00
1.93
CAP Cþ
Table 4 . The mean of CTDIvol , DLP , and effective dose for Abdomen C� , Abdomen Cþ , and AP Cþ .
Abdomen C�
Abdomen Cþ
AP
Parameters
Mean
Range
SD
Mean
Range
SD
Mean
Range
SD
CTDI vol
7,25
2,26 – 12,81
2.68
7,99
4,06 – 10,63
2.68
6,31
2,26 – 9,96
1.98
DLP
405,11
99,04 – 748,14
171.18
438,95
215,37 – 614,68
171.18
314,45
75,6 – 627,2
132.95
Effective dose
6,24
1,53 – 11,52
2.64
6,76
3,32 – 9,47
2.57
5,07
1,16 – 9,66
2.05
Fig . 1 . Distribution of CTDvol values during brain CT examinations . Fig . 2 . Distribution of DLP values during brain CT examinations .
CT-exams . The slice thickness is equal to 1,25 for all CTexams .
Tables 3 and 4 show that the mean CTDI vol varies from 6,31 mGy for AP with contrast media to 53,80 mGy for brain without contrast media . The mean DLP varies from 314,45 mGy . cm for AP with contrast media to 1131,44 mGy . cm for the brain without contrast media . The mean effective dose varied from 2,37 mSv for the brain without contrast media to 7,13 mSv for CAP with contrast media .
According to Figures 1 and 2 the CTDIvol values for brain CT vary from 20.73 to 78.04 mGy with a mean value of 53.80 mGy . DLP values vary from 607.07 to 1669.28 mGy . cm with a mean value of 1131.44 mGy . cm .
Table 5 shows that the DRL in terms of CTDI vol varies from 8,14 mGy for AP with contrast media to 64,45 mGy for the brain without contrast media . The DRL in terms of DLP varies from 417 mGy . cm for the chest without contrast media to 1405,77 for the brain without contrast media .
Table 6 shows that there was no association between patient gender and effective dose for all CT examinations . However , there is an association between patient age and the effective dose for brain CT without contrast media and chest CT without contrast media . In addition , there was an association between BMI and the effective dose of AP CT with contrast media and CAP CT with contrast media .
4 Discussion
The results show that , in terms of exposure parameters , the tube voltage is invariable for all examinations with a fixed value of 120 kV . In addition , the voltage does not change according to whether or not a contrast media is administered .