M. Zhou et al.: Radioprotection 2025, 60( 4), 373 – 381 377 Fig. 3. Comparison of two exposure systems and real case based on the induced voltage quantities for unipolar sensing and bipolar sensing. Table 1. Information of devices under test.
Device
Model
Lead Type
Tip-ring
Sensitivity
Sensitivity
Sensitivity
Tag
Spacing
Max.
Nom.
3rd
PM-I
Advisa SR
Active fixation
23 mm
0.45mV
Uni: 2.8 mV Bi: 0.9 mV
4 mV
PM-II
Adapta
Active fixation
10 mm
1 mV
2.8 mV
4 mV
ICD-I
Evera XT VR
Active fixation
10 mm
0.15 mV
0.3 mV
0.6 mV
ICD-II
Secura DR
Passive fixation
28 mm
0.15 mV
0.3 mV
0.6 mV
Table 2. Information of field sources.
Source
Directive and Standards
Exposure
Tag
Limit
Source-I
ICNIRP Reference Level( Public) / IEEE C95.1-2019( Unrestricted)
5 kV / m
Source-II
2013 / 35 / E. U. Low Action Level / ICNIRP Reference level( Occupational)
10 kV / m
Source-III
2013 / 35 / EU High Action Level / IEEE C95.1-2019( Restricted)
20 kV / m