DEPLOYMENT GUIDE
STEP 5 (Optional): ROTATE THE GUN BODY
If desired, push forward on the blue Knob
Lock to allow free rotation of the Gun Body.
The Gun Body can now rotate to the desired
deployment position for surgeon comfort
without disturbing Implant alignment in the
tunnel. Push back on the Knob Lock to
re-engage with the Gun Knob, preventing
further rotation of the Gun Body.
STEP 6: DEPLOY THE IMPLANT
To release the Deployment Trigger, squeeze the Deployment Trigger while pulling down on the blue Trigger Lock. While
maintaining slight forward pressure on the Gun, squeeze the Trigger to advance the Bullet into the Sheath. Significant
resistance to squeezing the Trigger indicates deployment. When the Trigger pull no longer exhibits resistance (i.e.
bottoms out), the Implant is fully deployed (approximately 2 trigger pulls).
Note: Be sure to apply slight forward pressure while pulling the Trigger to ensure that the Implant remains seated in the
tunnel.