END CAP SCREWS
•
• Headless PA Screws
• Length: 65 mm - 110 mm
available in 5 mm increments
• 3.5 mm Hex Head
Total Length: 6.85 mm
Head Length: 2.5 mm
Headed Cortical Screws
• Length: 20 mm - 60 mm
available in 5 mm increments
• 3.5 mm Hex Head
20 - 30.5 mm
Varies with length
Diameter
5 mm
Length 20 mm - 60 mm
Length 65 mm - 110 mm
Hex: 3 mm
Major Diameter
7 mm
Minor Diameter
5 mm
Major Diameter
5 mm
Minor Diameter
3.85 mm
ACCESSORY INSTRUMENTATION
The DynaFrame ™ CF Targeting System is made of high-strength, rigid carbon fiber polyether ether ketone (PEEK)
to provide accurate drill targeting and placement of screws along with excellent visibility under fluoroscopy.
Frame Operational Features
1. Cam Lever: Pull down to stretch the Compressive Element
2. Stop Wheel: Turn to set the desired length of Compressive Element stretch (0 - 6 mm)
3. Manual Compression Knob: Rotate clockwise to apply external compression
4. Deployment Base: Use lasermarks to determine amount of manual compression applied (0 - 6 mm)
5. Compression Tubes: Contain slots to aid in targeting medial-lateral screws
6. PA Targeting Arm: Attach to Compression Tubes to provide accurate placement of Headless PA Screw
7. Outer Tube Brace: Provides additional rigidity to the Targeting Frame
8. Step Numbers: Indicates order of steps involving the Targeting Frame
4
1
7
3
8
5
2
6
7