DynaNail® TTC Fusion System– Surgical Technique Guide DynaNail® TTC Fusion System– Surgical Technique Gu | Page 30

14 Proximal Tibial Screw Placement Instruments Used: 1. 8 mm Guide Sleeve (9) 2. Trocar Obturator (6) 3. 4 mm Drill Guide (8) 4. 4 mm Drill (23) 5. 3.5 mm Hex Driver Shaft (26) 6. Blue Handle Ratchet Driver (31) 7. Cortical Screw Thread the Trocar into a third 8 mm Guide Sleeve. Place the 8 mm Guide Sleeve and Trocar Assembly in the most proximal hole on the medial side of the Targeting Frame. Make an incision at the Trocar tip and advance down to bone. Leaving the Guide Sleeve in place, replace the Trocar with the 4 mm Drill Guide (indicated with a green stripe). Drill using a fresh 4 mm Drill, determine appropriate Cortical Screw length, and insert Cortical Screw as described in Step 12. Remove the 8 mm Guide Sleeve from the Targeting Frame. 30