DynaNail® TTC Fusion System– Surgical Technique Guide DynaNail® TTC Fusion System– Surgical Technique Gu | Page 20

10 Stretch Compressive Element Compressive Element visible Instruments Used: 1. Targeting Frame (30) Stretch the Compressive Element by pulling the lever into the full downward position (A). Use the Stop Wheel to set the desired amount of compression (up to 6 mm) from the Compressive Element (B, C). For example, setting the Stop Wheel to 6 signifies the Compressive Element is stretched by 6 mm and can therefore unload and maintain compression for up to 6 mm of bone resorption and/or settling. Flip the lever back up into its relaxed position (D). Note the DynaNail’s calcaneal screw holes have translated distally and are now in their final position. Confirm desired location of screw holes in the calcaneus under fluoroscopy before continuing. D A B 20 C L-M Screw Hole translated distal into calcaneus