3. Drill the second hole as described above. A second Locator Pin may be used for confirmation of depth and/or to maintain position prior to implanting the device. 4. Remove the Universal Drill Guide and Locator Pins. 5. When ready to introduce the implant to the prepared surgical site, remove the Inserter loaded with the DynaClip Implant from the sterile package. 6. Insert the DynaClip Implant into the drilled holes. 7. Pull up on the sleeve of the Inserter and slide the Inserter in the direction indicated by the arrow to disconnect the DynaClip Implant. 8. Seat the DynaClip Implant by using either manual pressure or the bottom metal tip of the Inserter to tamp the proximal end of the DynaClip Implant. 9. Radiographically, verify that arms have compressed or converged. 13. HOW SUPPLIED The DynaClip Bone Fixation System is provided sterile for single use only. Carefully inspect sterile packaging for damage prior to use. If the sterile packaging is found to be damaged or open, do not use the device or attempt to resterilize. Call your MedShape sales representative or MedShape Customer Service for a replacement. 14. STORAGE Store the DynaClip Bone Fixation System in a dry place at room temperature (20°C to 25°C).