4
SURGICAL TECHNIQUE
ARTHROSCOPIC BICEPS
Gun Body Rotation
5a
5b
If desired, push forward on the blue Knob Lock to allow free rotation of the Gun Body. The Gun Body
can now rotate to the desired deployment position for surgeon comfort without disturbing Implant
alignment in the tunnel. Push back on the Knob Lock to re-engage with the Gun Knob, preventing
further rotation of the Gun Body.
Implant Deployment
6a
Squeeze the Deployment Trigger while pulling
down on the blue Trigger Lock to release the
Deployment Trigger.
6b
While maintaining slight forward pressure on
the Gun, squeeze the Trigger to advance the
Bullet into the Sheath. When the Trigger no
longer exhibits resistance, the Implant is fully
deployed (approximately two pulls of the
Trigger).
Pearl #5. Be sure to apply slight forward pressure while pulling the Trigger to ensure that the Implant remains seated in
the tunnel.